Air Bubble Detector DYP-L01
L01 uses ultrasonic technology for bubble detection, which can correctly identify whether there are bubbles in any type of liquid flow.
Air Bubble Detector
Module : DYP-L01
The L01 series bubble sensor module is a module that uses ultrasonic sensing technology for detection.
It uses high-quality components, the product is stable and reliable,and has a long service life. This product is used for bubble detection in infusion pump products, automatic infusion alarms, and real-time monitoring of bubbles in other equipment and equipment pipelines.
Here comes the detailed specifications of DYP-L01 :
Item Description | Specification | Unit | Remark |
Operating voltage | 3.3~24 | V | DC |
Peak current | ≤25 | mA | |
Average operating current | ≤15 | mA | |
Response time | 0.2~100 | ms | |
Tube diameter | 3.5~4.5 | mm | Infusion tube |
Minimum bubble volume | 10 | uL | |
Duration | 2 |
Remarks:
(1)At room temperature, the power supply is 5V, and the typical data obtained by the test.
(2)Typical data obtained by testing with a 3.5mm diameter infusion tube under normal temperature, power supply 5V, measured medium is normal saline.
Rated Environmental Condition
Item | Minimum value | Typical value | Max value | Unit |
Storage Temp | -25 | 25 | 60 | ℃ |
Storage Humidity | 65% | 90% | RH | |
Operating Temp | 0 | 25 | 45 | ℃ |
Operating Humidity | 65% | 80% | RH |
Rated Electrical Condition
Item | Minimum value | Typical value | Max value | Unit | Remark |
Operating voltage | 3.2 | 5.0 | 24 | V | Peak value |
Peak current | 18 | 25 | mA | Peak value | |
Input Ripple | 50 | mV | Peak value | ||
Input Noise | 100 | mV | Peak value |
Product Features
Scope of Application
(The tested media include sterilized water,5%sodium bicarbonate,compound sodium chloride,10% concentrated sodium chloride,0.9%sodium chloride,glucose sodium chloride,5%-50%concentration glucose,etc.)
Model Selection
The output format of this series of ranging modules is divided into 5 formats. If there are special requirements that need to modify the response time, angle, communication agreement etc., you need to communicate with our FAE engineers during purchasing.
| No. | Output Interface | Model No. |
| L01 Series | GND-VCC Switch Positive output | DYP-L012MPW-V1.0 |
| VCC-GND Switch Negative output | DYP-L012MNW-V1.0 | |
| NPN output | DYP-L012MN1W-V1.0 | |
| TTL high level output | DYP-L012MGW-V1.0 | |
| TTL low level output | DYP-L012MDW-V1.0 |
Notice
(1)Please pay attention to the structural tolerances when designing. Unreasonable structural design may cause temporary abnormalities in module functions.
(2)Please pay attention to the evaluation of electromagnetic compatibility when designing. Unreasonable system design may cause malfunction of the module.
(3)When the boundary application of the product limit parameter is involved, you can contact after sale service dept.to confirm the relevant precautions.